Falcon 6/300R |
| MSE were one of the first manufacturers to introduce high capacity bench top centrifuges with induction drive. This tradition for innovation has continued with the introduction of the new Falcon range of large capacity bench top centrifuges. Both refrigerated and non–refrigerated models take a full 3 litre load – the refrigerated versionhaving a much increased ‘g’ force capability. These centrifuges represent excellent valuefor money in terms of their performance, ease of use, safety and reliability.
- Intelligent Microprocessor control
- Inverter controlled brushless motor
- Automatic Rotor Identification System
- Set and run in RCF or RPM
- High capacity Swing–out Rotor 4 x 750ml
- Up to 5,600 rpm with High Speed Rotor
- Key Lock facility
- Instant RCF reading
- 49 Program Memory
- Green easy read LED Display
- Overspeed protection system
- Full lid interlock
- Imbalance detection system
- Self diagnostics
|
|
- Status indicator
- CFC–Free refrigerant gas and insulation
- Fast efficient cooling
- Selectable temperature range
- Precool facility
- Simple and easy to use
- Easy to service
- 10 acceleration rates
- 10 brake rates
- Wide range of accessories
- All accessories fully autoclavable
- Sealed buckets for safety from aerosols
- Designed to comply with BS EN 61010–2–020: 1995
| - Automatic Rotor Identification System
- Both Falcon centrifuges incorporate the well known MSE automatic rotor identification system. This important feature ensures that the centrifuge recognises the rotor in use and prevents it from being used beyond its maximum speed. Once the centrifuge has identified the rotor it can also work out the ‘g’ force at any set speed. This means that the centrifuge can convert rpm to ‘g’ and vice versa once a run has commenced.
- Microprocessor Controlled Induction Drive
- Brushless induction drive motors have become a familiar feature in bench top centrifuges since they were first introduced by MSE in the early 1980’s. Naturally both Falcons follow this tradition. Combined with microprocessor control, each version exhibits extremely accurate run to run reproducibility – an important feature for users where accountability for analysis and production is becoming an increased necessity.
- Fully Programmable
- Operation of both units is fully programmable with a memory capability of up to 49 different programs. This includes all parameters such as speed (or ‘g’ force), time, acceleration rate, brake rate and temperature control.
- Safety Features
- MSE have always prided themselves on customer safety and the Falcons prove to be no exception. Both centrifuges have a 7mm steel guard ring so that in the unlikely event that there should be a disruption, no debris can escape from the centrifuge. An imbalance sensor aborts a run if a significant imbalance is detected and a lid interlock prevents the lid from being opened when the rotor is in motion.
- Wide Range of Accessories Available
- Most centrifuge work can be accomplished using the 4 x 750ml windshielded rotors, as these are available with a wide range of adaptors that fit into the 750ml buckets. Microtitre carriers and special buckets for 250ml tissue culture bottles are available on the 4 x 750ml non–windshielded rotor only. The 8 x 200ml swing–out rotor and various angle rotors complete the full list of accessories making this one of the most versatile centrifuges on the market. Please see rotor tables for details.
Specifications
| Specifications |
| Max. Speed Swing–Out Rotor |
4,700 rpm |
| Max. RCF Swing–Out Rotor |
4,964 x g |
| Max. Speed Titanium Swing–Out Rotor |
5,600 rpm |
| Max. RCF Titanium Swing–Out Rotor |
7,047 x g |
| Max. Speed Angle Rotor |
6,000 rpm |
| Max. RCF Angle Rotor |
6,030 x g |
| Max. Bottle Size |
750ml |
| Max. Capacity |
3,000ml |
| Speed Control Range |
200–6,000 rpm |
| Speed Control Accuracy |
1 rpm |
| Speed Display |
±10 rpm |
| Temperature Range |
−9°C to +40°C |
| Temperature Control |
±2°C |
| Time Set Range |
| 0–59 mins 59 secs |
| 0–99 hours 59 mins |
| Time Hold | |
| Time Accuracy |
0.1 secs |
| Dimensions |
| Height |
433mm |
| Width |
826mm |
| Depth |
600mm |
| Weight |
146kg |
Adaptors
| Rotors and Associated Buckets/Microplate Carriers (Refrigerated) |
| Swing–Out Rotors |
| Cat. No. |
Description |
Buckets for Rotor |
Microplate Carriers |
Buckets for 250ml Conical Bottles |
| 43124–1298 |
750ml Non–Windshield Rotor Max. Speed 2,500 rpm (1,317/1,401 x g) |
43551–1235 Set of 4 x 750ml Sealed Buckets |
41159–1900 Set of 4 Microplate Carriers. Holds 3 Plates per Carrier. Max.Speed 2,500 rpm (1,329 x g) |
43551–130 Set of 4 Buckets for 250ml Conical Tissue Culture Bottles |
| 43124–7080 |
750ml Windshield Rotor Max. Speed 3,600 rpm (2,828/3,000 x g) |
43551–1235 Set of 4 x 750ml Sealed Buckets |
Not Possible with this Rotor |
Not Possible with this Rotor |
| 43124–756 |
750ml Windshield Rotor Max. Speed 4,700 rpm (4,964 x g) |
43551–141 Set of 4 x 750ml Sealed Buckets |
Not Possible with this Rotor |
Not Possible with this Rotor |
| 43124–757 |
750ml Windshield Rotor Max. Speed 5,600 rpm (7,047 x g) |
43551–142 Set of 4 x 750ml Sealed Buckets |
Not Possible with this Rotor |
Not Possible with this Rotor |
| Select the required adaptors to fit your tubes from the tables below |
| Angle Rotors – See Angle Rotor Table Below | In addition to the rotors details in the two tables below, the following rotors are available: - Windshielded Rotor 43124–7080 (3,600 rpm) Max. RCF range 2,823 – 3,000 x 'g'
- Non–Windshield Rotor 43124–1298 (2,500 rpm) Max. RCF range 1,317 – 1,401 x 'g'
Adaptors for use with 750ml Sealed Buckets Windshielded Rotor 43124–756 (4,700 rpm) |
| Cat. No. |
Adaptor |
Colour |
Rotor Capacity |
Max. Dia. |
Max. Length |
Min. Length |
Max. RPM |
Max. RCF |
| 43155–1322 |
37 x 5ml |
Blue |
148 |
12.0mm |
125mm |
50mm |
4,700 |
4,655 |
| 43155–1335 |
28 x 7ml |
Orange |
112 |
13.0mm |
125mm |
48mm |
4,700 |
4,655 |
| 43155–134A |
19 x 15ml |
Green |
76 |
17.0mm |
125mm |
63mm |
4,700 |
4,779 |
| 43155–1354 |
5 x 50ml |
Red |
20 |
32.0mm |
125mm |
77mm |
4,700 |
4,879 |
| 43155–1369 |
7 x 30ml Flat |
Yellow |
28 |
28.5mm |
125mm |
67mm |
4,700 |
4,655 |
| 43155–152 |
6 x 50ml Falcon |
Beige |
24 |
32.0mm |
125mm |
78mm |
4,700 |
4,879 |
| 43155–153 |
12 x 15ml Falcon |
Maroon |
48 |
17.0mm |
125mm |
75mm |
4,700 |
4,779 |
| 43551–1269 |
Blood Bag Liner |
White |
4 |
98.0mm |
— |
— |
4,700 |
4,964 |
| 43551–124A |
750ml PP Bottle |
White |
4 |
98.0mm |
— |
— |
4,700 |
4,964 |
Adaptors for use with 750ml Sealed Buckets Windshield Rotor 43124–757 (5,600 rpm) |
| Cat. No. |
Adaptor |
Colour |
Rotor Capacity |
Max. Dia. |
Max. Length |
Min. Length |
Max. RPM |
Max. RCF |
| 43155–1322 |
37 x 5ml |
Blue |
148 |
12.0mm |
125mm |
50mm |
5,600 |
6,609 |
| 43155–1335 |
28 x 7ml |
Orange |
112 |
13.0mm |
125mm |
48mm |
5,600 |
6,609 |
| 43155–134A |
19 x 15ml |
Green |
76 |
17.0mm |
125mm |
63mm |
5,600 |
6,784 |
| 43155–1354 |
5 x 50ml |
Red |
20 |
32.0mm |
125mm |
77mm |
5,600 |
6,925 |
| 43155–1369 |
7 x 30ml Flat |
Yellow |
28 |
28.5mm |
125mm |
67mm |
5,600 |
6,609 |
| 43155–152 |
6 x 50ml Falcon |
Beige |
24 |
32.0mm |
125mm |
78mm |
5,600 |
6,925 |
| 43155–153 |
12 x 15ml Falcon |
Maroon |
48 |
17.0mm |
125mm |
75mm |
5,600 |
6,784 |
| 43551–1269 |
Blood Bag Liner |
White |
4 |
98.0mm |
— |
— |
5,600 |
7,047 |
| 43551–124A |
750ml PP Bottle |
White |
4 |
98.0mm |
— |
— |
5,600 |
7,047 |
| Angle Rotors |
| Rotor |
Bucket |
Adaptor |
Capacity |
Max. Diameter |
Max. Length |
Max. Speed |
Max. RCF |
| 43117–605A (6 Place Universal) |
34411–9120 (Bucket, S/Steel, 100ml) |
34142–1057 (Rubber Cushion) |
6 x 100ml |
42mm |
125mm |
6,000 |
5,524 |
| 34144–3087 |
6 x 50ml |
29mm |
120mm |
6,000 |
5,108 |
| 34151–302A |
12 x 15ml |
17mm |
118mm |
6,000 |
5,524 |
| 34153–302 |
30 x 7ml |
13mm |
118mm |
6,000 |
5,311 |
| 43114–609A (16 Place Universal) |
Not Required |
Not Required |
16 x 15ml |
29mm |
110mm |
6,000 |
5,920 |
| 34144–3072 |
16 x 50ml |
17.5mm |
110mm |
6,000 |
5,720 |
| 43114–6085 (24 Place Universal) |
Not Required |
Not Required |
24 x 15ml |
17mm |
108mm |
6,000 |
6,030 |
| 43117–610 (8 Place Sealed Rotor) |
43156–603 (Bucket, Sealed, Aluminium) |
Not Required |
8 x 50ml |
29mm |
120mm |
6,000 |
5,635 |
| 34142–110 (50ml Conical Tube Adaptor) |
8 x 50ml |
29mm |
120mm |
6,000 |
5,635 |
Ordering Information
| Model |
Description |
| MSB300.CR3.K |
Falcon 6/300R 220–230V 50Hz |
| MSB300.CR3.1 |
Falcon 6/300R 120V 60Hz | Popular Systems
| Falcon 6/300R with 4 x 750ml Swing–Out Rotor (Max. Speed 3,660 rpm) |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43124–7080 |
Windshield Swing–Out Rotor |
1 |
| 43551–1235 |
Sealed Buckets, 750ml (x4) |
1 |
| Adaptors |
Choose the relevant adaptors from above to suit your tubes (sold in sets of 2) |
2 |
| |
| Falcon 6/300R with 4 x 750ml Swing–Out Rotor (Max. Speed 4,700 rpm) |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43124–756 |
Windshield Rotor, 4,700 rpm, 4 x 750ml |
1 |
| 43551–141 |
Sealed Buckets, 750ml (x4) |
1 |
| Adaptors |
Choose the relevant adaptors from above to suit your tubes (sold in sets of 2) |
2 |
| |
| Falcon 6/300R with 4 x 750ml Swing–Out Rotor (Max. Speed 5,600 rpm) |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43124–757 |
Windshield Rotor, 5,600 rpm, 4 x 750ml |
1 |
| 43551–142 |
Sealed Buckets, 750ml (x4) |
1 |
| Adaptors |
Choose the relevant adaptors from above to suit your tubes (sold in sets of 2) |
2 |
| |
| Falcon 6/300R with 4 x 750ml Swing–Out Rotor (Max. Speed 2,500 rpm) to Spin Corning Conical Tissue Culture Bottles |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43124–1298 |
Swing–Out Rotor (Non–Windshield) |
1 |
| 43551–130 |
Sealed Cups, Corning T/Culture (x4) |
1 |
| |
| Falcon 6/300R with 4 x 750ml Swing–Out Rotor (Max. Speed 2,500 rpm) to Spin Up To 12 Microtitration Plates |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43124–1298 |
Swing–Out Rotor (Non–Windshield) |
1 |
| 41159–1900 |
Set of 4 Microtitre Carriers |
1 |
| |
| Falcon 6/300R with 24 x 15ml Angle Rotor |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43114–6085 |
Angle Rotor, 24 Place |
1 |
| |
| Falcon 6/300R with 16 x 50ml Angle Rotor (not for 50ml Falcon Tubes) |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43114–609A |
Angle Rotor, 16 Place |
1 |
| 43156–602 |
Buckets, Matched, 50ml (x4) |
4 |
| |
| Falcon 6/300R with 6 x 100ml Angle Rotor |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43117–605A |
Universal Angle Rotor, 6 Place |
1 |
| 34411–9120 |
Buckets, Matched, 100ml (x4) |
2 |
| |
| Falcon 6/300R with 8 x 50ml Sealed Angle Rotor (can be used with 50ml Falcon Tubes) |
| Cat. No. |
Description |
Quantity |
| MSB300.CR3… |
Order centrifuge to suit local voltage (see above) |
1 |
| 43117–610 |
Sealed Angle Rotor, 8 x 50ml |
1 |
| 43156–603 |
Sealed Buckets, Aluminium, 50ml (x4) |
2 |
|